4. Installation
DEPOLOX® Pool E flow cell
26
EN
9 Screw the flow distributor cap back on.
10 Open the shut-off ball valve on the sam-
ple water inlet and outlet. The cell body
fills with sample water again.
11 Refit the housing cover of the flow cell.
12 After 2 to 3 hours running-in time, per-
form a chlorine calibration. If necessary,
repeat the chlorine calibration after 24
hours (see chapter “Calibration“ in the
instruction manual “700 P Electronics
module“).
4.13 Plugging sensors into the
DFMe electronics module
1 Insert the multi-sensor (A) in the flow
control valve.
2 Connect sensor cable Cl
2
free (B) to the
chlorine sensor.
3 Insert or screw the LED glow stick (C) into
the mounting hole in the cell body cover.
4 Connect the pH- and ORP (Redox) sensor
cable plug (D, E) with the pH and ORP
(Redox) single-rod measuring chain.
5 Plug the CAN connection cable into the
CAN connection socket (L) and with con-
nect with 700 P electronics module (CAN
socket).
Fig. 8
Cross-section, DFMe electronics
module
A Multi-sensor
B
Sensor cable Cl
2
free
C
LED glow stick
D pH sensor cable
E
ORP (Redox) sensor cable
F
Housing DFMe electronics module
G Mounting hook
H Sensor measuring module pH
I
Sensor measuring module ORP (Redox)
J
DFMe motherboard with measurement in-
put Cl
2
K
CAN extension socket
L
CAN connection socket for 700 P electro-
nics module
NOTICE
An initial rotating air bubble at the bottom
of the cell body does not affect the measu-
rement.
A
B
C
H
I
F
J
K
L
D
E
H
I
G