E2660-CO-NO2

Rev 06.04.2021

Product description

Dual Gas Detector-Transmitter E2660R-CO-NO2 belongs to the PluraSens® family of
multifunctional measurement instruments. The device is intended for simultaneous
detection of carbon monoxide and nitrogen dioxide. It is a convenient solution for
underground parking. The instrument utilizes electrochemical gas sensors with excellent
repeatability, stability, and long lifetime.

Two analog outputs and RS485 digital interface with industry-standard Modbus RTU
protocol can be used to connect the device to safety or building automation systems.

The following features are available as an option:

Two relays RE1 and RE2 with closing contacts. Relay outputs can be used to
switch alarm sirens, ventilation fans, shut-off valves, or other actuators,

24 VAC or 230 VAC as options for power supply module,

Remote probe.

The version of your detector is marked on the package. If the symbol

is marked on the

equipment, consult the documentation for further information.

Safety requirements

Misuse will impair the protection of the product. Always adhere to the safety provisions
applicable in the country of use.

Do not perform any maintenance operation with the power on. Do not let water or foreign
objects inside the device.

Removal of the PCB from the enclosure voids the warranty. Do not touch the electronic
components directly, as they are sensitive to static electricity.

Connection diagrams can be found in the electrical connections section. The device might
not perform correctly or be damaged if the wrong power supply is connected.

External circuits connected to the equipment should have sufficient insulation rating
according to the environmental conditions and equipment power.

A disconnecting device that is marked as such and easily accessible should be included in
the installation of this product.

7

Summary of Contents for PluraSens E2660-CO-NO2

Page 1: ...PluraSens Dual Gas Detector Transmitter E2660 CO NO2 User Manual...

Page 2: ...dimensions 9 Sensor probe handling 10 Gas sensor replacement procedures 10 Electrical connections 11 Operation 13 Maintenance 13 Calibration 13 Delivery set 14 Order code for E2660 CO NO2 options 14...

Page 3: ...losive limit UEL vol in air 74 Odor Odorless Hazards Highly toxic Mild poisoning causes lightheadedness confusion headache dizziness and flu like effects Larger exposures can lead to toxicity of the C...

Page 4: ...te Irritates the skin eyes and respiratory tract Exposure to levels above 100 ppm can cause death due to asphyxiation from fluid in the lungs There are often no symptoms at the time of exposure other...

Page 5: ...scale 0 1000 0 200 0 100 Operating temperature 20 50 C 30 50 C 20 50 C Operating conditions 15 90 RH atmospheric pressure 10 non ATEX rated areas Non aggressive atmosphere Cross sensitivity Acetylene...

Page 6: ...19 EN 61000 6 2 2019 EMC Immunity EN 60079 29 1 2016 EN 60079 29 2 2015 and EN 60079 29 3 2014 Relay option Output relays 2 SPST relays closing contact 250 VAC 30 VDC 5 A max Alarm set points RE1 set...

Page 7: ...wer supply module Remote probe The version of your detector is marked on the package If the symbol is marked on the equipment consult the documentation for further information Safety requirements Misu...

Page 8: ...etry direction and velocity of airflows etc Carbon monoxide has practically the same density as air Nitrogen dioxide is 1 6 times heavier than air but in the underground parking it rises to the ceilin...

Page 9: ...E2660 CO NO2 Rev 06 04 2021 For installing the device attach the device to the wall using provided screws through the side mounting holes or key slots see mountains dimensions Mounting dimensions 9...

Page 10: ...move the old filter Place a new filter into the nut and tighten it again NOTE Never stab or press the filter near its center where the sensor is located since this may damage the sensor Do not remove...

Page 11: ...ply and of the external devices Attach the power cable to the device without turning it on Using the connection diagram below connect the analog outputs and digital interface terminals to the relevant...

Page 12: ...85 interface should not exceed 1200 m Both analog outputs can be independently changed between 4 20 mA and 0 10 V type using jumpers J1 OUT1 and J2 OUT2 By closing pins on a specific jumper the relate...

Page 13: ...ernal devices connected the power LED green on and the control LED red is constantly lit Make certain that the cable glands are properly tightened to ensure the conformity to IP65 protection class Pla...

Page 14: ...lamp for remote probe versions Order code for E2660 CO NO2 options E2660 options Order code Remote probe 3 m cable E2660 CO NO2 RP33 3 Remote probe 10 m cable E2660 CO NO2 RP33 10 Integrated 90 265 V...

Page 15: ...meter Flexible setting of analog output scales for each output Output zero and slope adjustment for calibration Free assignment of each of two relays to the chosen parameter Several relay control logi...

Page 16: ...register Error codes 01 Illegal function 02 Illegal data address 03 Illegal data value 04 Slave device failure details of last error 04 can be read from register 0x0008 Modbus holding registers Regist...

Page 17: ...ent for gas data 1 65535 512 0x00A7 167 40168 RW The change rate limit for gas data ppm gas unit s 1 32000 0 no limit 0 0x00A8 168 40169 RW Integrating filter time constant s 1 32000 seconds 0 no filt...

Page 18: ...8 RW Minimal on off time for relay RE1 0 1000 s 0 0x00DA 218 40219 RW Minimal on off time for relay RE2 0 1000 s 0 Addr Reg MHR RW Description Supported values dec Default 0x00DB 219 40220 RW Control...

Page 19: ...nt turn signaling off on analog output1 bit 3 0 1 in case the sensor is absent turn on signaling with low current high current on analog output1 if bit 2 0 this bit will be ignored bit 4 0 1 in case o...

Page 20: ...DC 32000 32000 ADC units 0 0x016E 366 40367 Slope adjustment for 2nd gas data 1 65535 0x016F 367 40368 The change rate limit for 2nd gas data gas unit s 1 32000 0 no limit 0 0x0170 368 40369 2nd Integ...

Page 21: ...Manufacturer will at its option either repair or replace a product that proves to be defective This warranty is void if the product has been operated in conditions outside ranges specified by the Man...