E2218
Rev 14.05.2021
normal operating -
continuously on/off depending
of bit[10]
0x0102
(4)0259
R
Measured temperature,
°C×100
signed integer, -4000...+12500
(-40.00...+125.00 °C)
-
0x0103
(4)0260
R
Measured humidity,
%RH×100
integer, 0...+10000
(0.00...100.00 %RH)
-
0x0104
(4)0261
R
Calculated dewpoint,
°C×100
signed integer, -8000...+10000
(-80.00...+100.,00 °C)
-
0x0105
(4)0262 RW
0% value for analog
output 1, °C / % RH / g/m
3
signed integer, -1000...+1000
(-1000...+1000 °C/%RH)
0
0x0106
(4)0263 RW
100% value for analog
output 1, °C / % RH / g/m
3
signed integer, -1000...+1000
(-1000...+1000 °C/%RH)
100
0x0107
(4)0264 RW
0% value for analog
output 2, °C / % RH / g/m
3
signed integer, -1000...+1000
(-1000...+1000 °C/%RH)
0
0x0108
(4)0265 RW
100% value for analog
output 2, °C / % RH / g/m
3
signed integer, -1000...+1000
(-1000...+1000 °C/%RH)
100
0x0119
(4)0282
R
Calculated absolute
humidity, g/m
3
×100
0...+32767 (0...327.67 g/m
3
)
-
* – The new value is applied after restart.
** – Broadcast slave ID 0 can be used to assign a new ID to the instrument with an
unknown ID. When addressing by ID 0 the device shall be the only Modbus instrument in
the network. The device will not respond to the Master command when addressed by ID 0.
*** – This value is dynamic and not kept in EEPROM after a restart.
16