Instruction Manual
CyberScan pH5500/ 5000
To take a measurement
1. Immerse the ion selective electrode in a known volume of the
sample solution.
2. Stir the sample.
3. Touch
KA
to take the first mV reading of sample. Wait for the
reading to stabilize (
STABLE
will appear).
OR
Touch
KA
to accept the current reading.
4. Add a known volume of standard solution to the sample.
5. Touch
KA
to take the second mV reading of sample. Wait for the
reading to stabilize (
STABLE
will appear).
OR
Touch
KA
to accept the current reading.
6. Use the keypad to enter the sample volume.
7. Touch
enter
to accept the value.
8. Use the keypad to enter the standard volume.
9. Touch
enter
to accept the value.
10. Use the keypad to enter the concentration of the standard
solution added to the sample.
11. Touch
enter
to accept the value.
12. Use the keypad to enter the concentration of the standard
solution added to the sample
13. Touch
enter
to accept the value. Your Known Addition result is
now displayed.
14. Touch
to send the data to memory if sample ID# has been
assigned or to print the data on the screen.
OR
Touch
OK
to return to the Measure screen for a new
measurement.
114