14
4.3
pH Calibration (without ATC)
When the CyberScan meter is not used with a temperature probe, it is necessary to calibrate
the temperature first prior to pH calibration.
1.
Calibrate for temperature as described in the procedure in Section 4.4.1 - Temperature
Calibration Without ATC Probe.
2.
Next, proceed with the pH calibration as described in the above section.
3.
The meter is now calibrated at a fixed temperature.
NOTE: To exit this program without confirming the temperature calibration value, DO NOT
press CON. Press CAL/MEAS instead.