Instruction Manual
CON 11/110
52
8.10.2 P9.2: Selection of Single or Multi Point Calibration
The single point calibration lets you have a single calibration factor in all the five
measuring ranges by just calibrating one point in either one of the five measuring
ranges. When used in an automatic or manual calibration mode the calibration factor
is applied to all the five ranges
The multi point calibration, on the other hand, gives you better calibration accuracy
by letting you calibrate in each of the five ranges. Whether it is used in an automatic
or manual calibration mode, the calibration factor would only
be applicable in the particular range where the calibration is
done.
1. Press SETUP key to enter Set Up menu.
2. Press the MI/
▲
or MR/
▼
keys until you view
parameter P9.0.
3. Press the HOLD/ENTER key three times to enter
parameter P9.2. The upper display will show “SPC”.
4. Press the HOLD/ENTER key to enter the single or
multi point calibration selection mode.
5. Press
the
MI/
▲
or MR/
▼
keys to select “YES” or
“NO”.
•
YES activates the single point calibration.
•
NO activates the multi point calibration.
6. Press the HOLD/ENTER key to confirm selection.
7. Press the CAL/MEAS key to return to measurement
mode.
MI
MR
Spc
SETUP
P 9.2
Spc
SETUP
Yes
Spc
SETUP
HOLD
ENTER
Figure 45: P9.2-
Setting Single or
Multi point
Calibration