Instruction Manual
CON 11/110
41
8.2.4 P1.4: Selection of TDS Conversion Factor
You can set the TDS conversion factor between 0.40 and
1.00; meter default is 0.50.
NOTE: Your meter has to be in the TDS measurement mode
before you can make the selection of the TDS Conversion
Factor in the SETUP menu.
Only from TDS measurement mode
1. Press SETUP key to enter Set Up menu.
2. Press the HOLD/ENTER keys until you view
parameter P1.4.
3. Press the HOLD/ENTER key again. The upper
display shows a value and the lower display shows
“tdS”.
4. Press the MI/
▲
or MR/
▼
keys to select your
calculated TDS conversion factor.
5. Press the HOLD/ENTER key to confirm selection
and to return to the subgroup menu.
6. Press the CAL/MEAS key to return to measurement
mode.
ppt
ppm
SETUP
p 1.4
ppt
ppm
SETUP
0.50
ppt
ppm
SETUP
0.75
MI
MR
HOLD
ENTER
Figure 34: P1.4 -Select
TDS Conversion
Factor