Instruction Manual
CON 11/110
36
8.1
Advanced SETUP Mode Overview
1. Press the SETUP key to enter Set Up menu.
2. Press
the
MI/
▲
or MR/
▼
keys to scroll through sub groups.
3. Press HOLD/ENTER key to enter a particular parameter.
SETUP
P 1.0
COF
P1.0: Unit Configuration
P1.1 Selection of READY and *Auto HOLD function (*CON
110
Only).
P1.2 Selection of °C or *°F (*CON 110 Only)
P1.3 Selection of Automatic or Manual Temperature
Compensation
P1.4 Selection of TDS Conversion Factor (This mode only
appears when SETUP Menu is chosen while in the
TDS measurement mode)
SETUP
CAL
P 2.0
P2.0: Viewing previous calibration data
P2.1 First range calibration point
P2.2 Second range calibration point
P2.3 Third range calibration point
P2.4 Fourth range calibration point
P2.5 Fifth range calibration point
SETUP
ELE
P 3.0
P3.0: Viewing electrode data
P3.1 Effective cell constant for first range
P3.2 Effective cell constant for second range
P3.3 Effective cell constant for third range
P3.4 Effective cell constant for fourth range
P3.5 Effective cell constant for fifth range
SETUP
P 4.0
A o
P4.0: Auto Off
P4.0 Selection
of
automatic power off function
SETUP
P 5.0
P5.0: Reset to factory defaults
P5.1 Resetting of all calibration and clear all
calibration
data
P5.2 Resetting of all User settings to factory
defaults
condition
SETUP
P 6.0
MEM
P6.0: Memory clear
P6.0 Clearance of all stored readings
SETUP
P 7.0
°C °F
P7.0: Temperature
P7.1 Setting of temperature coefficient
P7.2 Setting of normalisation temperature