10.4 Firmware
Upgrade
10.4.1
Firmware Upgrade: PC software
In order to upgrade the firmware of the instrument proceed as follows:
1 - Install the ST167Flash program from the CD.
2 - Connect the infrared interface adapter to the serial port of the PC with the supplied cable (note back the serial port
used: COM1 , COM2 or other).
3 - Position the interface near the right side of the instrument.
4 - Run the ST167Flash program you have installed.
5 - Select the same serial port that you noted back.
6 - Select the Baud rate to 115200 Baud (if the PC owns an UART 16550 type), otherwise select a lower Baud rate.
7 - Press the three dotted button in order to load the firmware file (.HFW extension) normally located on a 3 1/2" diskette.
A version number of the firmware being discarded is shown in the dedicated field.
8 - With instrument OFF press and hold the F1 key while pressing the ON/OFF key in order to set the instrument in
firmware upgrade status.
9 - Click on the 'Program' button to start the upgrading phase and wait until a termination message appears.
NOTE:
I
F DATA TRANSFER FAILS
,
PLEASE TURN OFF THE INSTRUMENT AND REPEAT THE PROCEDURE FROM STEP
8.
10.4.2 Firmware
Upgrade
between
two Uni units: bootloader
The firmware of the UniGas 3000+ can be replaced by the higher version of another UniGas 3000+ if needed. Simply
connect together the two instruments. To launch the procedure switch on the two instruments after you have connected
them together.
•
Switch on the instrument and go to the analysis page
•
Press the [
HELP
] key on the keyboard of the instrument where you want load the new firmware: the following page
will appear:
HELP 15:44 20/06/06
HEADER PRINTER
SENSOR DIAGNOSIS
Batt. 100%
BootLoader
ESCAPE
–
SERVICE
•
Use the [
S
] and [
T
] keys to select "
BootLoader
" and press [
ENTER
]
•
At the end of the firmware transferring press [
ESCAPE
] and switch off the instrument to exit the configuration mode.