HA250385
Issue 2 Mar 98
Section 2
Page 2 - 6
MULTIPOINT CIRCULAR-CHART RECORDER INSTALLATION AND OPERATION MANUAL
2.5.2 Entering Configuration (Cont.)
The password is set to '10' at the factory. The pass-
word can be edited in Instrument Configration.
To enter '10', press the cursor key three times, then the
up arrow and 'enter' keys.
2.5.3 Channel configuration
Operate the Page up key twice, to call the top level
channel configuration page.
Use the up arrow key to change the channel number to
'2' and operate the 'Enter' key.
Use the Enter key again to enter 'Range' configuration.
Operate the up arrow key to change 'Off' to 'T/C'
(Thermocouple). Note the other input type choices.
The 'E' LED illuminates to remind you that you have
made a change which has not yet been entered into the
data base.
Operate the Page up key to call the Range low page.
The low range is 0 as required, so operate the page key
again to call the Range High page
Enter the value 1000 as follows:
1.
Operate the up arrow key until '1' appears in the
display.
2.
Operate the cursor key, and repeat step 1, but stop
when '0' appears in the display.
3.
Repeat step 2.
4.
Repeat step 2.
5
Repeat step 2 but stop when the decimal point
appears.
Note: If you do not enter the decimal point, the re-
corder will interpret the entry as 100000
Password 00000
Password 00000
Password 00000
Password 00000
Password 00010
Conf: Instrument
Conf:Chart
Conf:Channel 1
Conf:Channel 2
Channel : Range
I/P Type Off
I/P type T/C
I/P Range Lo 0.00
I/P Range Hi 0.00
I/P Range Hi 1000.0
'E'
'E'
'E'
'E'