27
Eurotek RBT 7000
Calibration of weight unit
Start the software and enter the sub-menu by click the “F8-Configuration”,
select the item of Weight calibration
Press start to continue.
Make sure there is no force applying to the sideslip plate, and then click “Zero” to
set the Zero point value for sideslip sensor.
Add load (Known Weights) directly to the tester bench (in appropriate measurement range
2000-5000 N)
,
verify the display value of the screen(Display value = Added Known Weight in kg).
If the display is exceed the error tolerance, please use the “Ratio”value to adjust the Display
value until the correct.