85
EVENT SUMMARY
Provides the current therapy times displaying 5 records divided into 2 pages:
Therapy start
Provides the date and time the device was switched on and
NEW THERAPY
mode was
selected on the start page.
Total therapy time
Provides the total time of the current therapy from the start of
NEW THERAPY.
ON time
Provides the total time of the current therapy in which the device was on without
alarms.
OFF time
Provides the total time of the current therapy in which the device was off and/or in
alarm (i.e. the therapy interruption time).
ALARM time
Provides the total time of the current therapy in which the device was in alarm (this time
is part of the
OFF Time
).
After viewing the desired pages, press
ESC
to return to the page that includes the events summary, events details
and download events functions.
EVENT DETAILS
Provides in detail with date and time all the events and parameters set from the start of the current therapy,
including all the alarms that have appeared and been cleared. The display is divided into several pages for easier
understanding.
After viewing the desired pages, press
ESC
to return to the page that includes the events summary, events details
and download events functions.
Note
In order to correctly read the alarms recorded by the data logger, refer to the
table below.