Operation
|
31
In spatial averaging, a series of measurements are taken during a
relatively short period of time (usually not more than two to three
minutes). At the end of the period, the average value is the answer.
The measurement is used to monitor compliance with standards,
based on the whole-body average maximum permissible exposures
(MPE). The average field is measured over the space occupied by the
subject’s body.
For more information about averaging see
Averaging
on page 38.
A
LARM
P
OINT
Alarm Point
allows the user to set a field level above which an alarm will
indicate/sound. The range is 0.00 to 1000 units. Audio must be set to
Alarm
for
an audible tone to sound when the alarm level is reached. If the alarm level is set
higher than the range of the probe, the alarm will not indicate/sound. Setting the
alarm level to 0.00 will turn the alarm off.
A
UDIO
Audio
is an audio annunciator that is provided to signal various functions on the
HI-2200. Options include
None
,
Alarm
,
Cadence
, and
Both
.
•
None
turns off all sounds excluding the startup tone.
•
Alarm
causes the unit to emit a tone when the alarm level set under
the Alarm Point option is exceeded.
•
Cadence
enables a pulsing tone to assist in measuring consistently
over a distance. This tone sounds every second during logging in
continuous or spatial log mode.
•
Both
enables the alarm point tone and the cadence tone at the same
time. In addition to the pulse tone, a continuous tone will sound when
the alarm level is reached
Summary of Contents for HI-2200
Page 5: ... v Appendix B Menu Tree 47 Appendix C EC Declaration of Conformity 53 ...
Page 6: ...vi This page intentionally left blank ...
Page 8: ...viii This page intentionally left blank ...
Page 20: ...20 Specifications This page intentionally left blank ...
Page 22: ...22 Assembly and Installation This page intentionally left blank ...
Page 46: ...46 Warranty This page intentionally left blank ...
Page 47: ...Menu Tree 47 Appendix B Menu Tree ...
Page 48: ...48 Menu Tree Units available are probe dependent ...
Page 49: ...Menu Tree 49 ...
Page 50: ...50 Menu Tree ...
Page 51: ...Menu Tree 51 ...
Page 52: ...52 Menu Tree This page intentionally left blank ...
Page 53: ...EC Declaration of Conformity 53 Appendix C EC Declaration of Conformity ...