power off.
Under PEAK test mode, firstly press
POWER
button for
exit from PEAK test mode, returning to general test mode,
then press
POWER
button for power off. That is, under PEAK
test mode, it needs to press
POWER
button twice for power
off.
Under data reference mode, firstly press
POWER
button
longer (over 3 seconds) for exit from data reference mode,
returning to general test mode, then press
POWER
button for
power off. During the exit from data reference, it will show
“End” symbol.
2
.
Gear Shift Choice
Only under test mode, it is available for gear choice.
Gear choice button
RANGE
is self-lock button, so it
needs to press a bit heavier.
Press
RANGE
button to choose 10A or 600A level, and
the relevant indicator light will light. If the current on test is
unknown, please choose 600A level firstly, and then make the
most suitable gear level according to the test result.
Under data reference mode, both gear shift indicator and
PEAK light are not bright.
3
.
General Test
High voltage, extremely dangerous! It must be
operated by qualified staff that has acquired
authority. The operator must strictly abide by the
safety rules. Otherwise, there is electric injury
danger, causing personal damage or casualty
accident.
Dangerous! It cannot be used to measure the
circuit over 23KV voltage. Otherwise, there is
electric injury danger, causing personal damage
or equipment damage.
Dangerous! It cannot be used to measure the
circuit over 600A voltage. Otherwise, there is
electric injury danger, causing personal damage
or equipment damage.