CONTROLLER DESCRIPTION 

The eSelect-CD1 is a compact system with ELECTRICAL CONDUCTIVITY (EC) controller. In many cases, conductivity is linked directly 

to the total dissolved solids (T.D.S.): conductivity provides an approximate value for the TDS concentration, usually to within ten-percent 

accuracy.  eSelect-CD1, along with the high quality performance in terms of measurements and functions thus satisfying most of the 

requirements for an instrument, it offers many features thus making it extremely versatile and easy to operate. Conductivity measurement are 

in micro Siemens features 4 K factor ranges (Cell Constant) suitable for following K: 

Controller covers ranges from costant K 1 up to 20.00 mS and even 100.00 mS (latter with graphite sensor electrodes) and costant K 5 

ranges 2.000 μS;   

BENEFIT: The above feature allow dealers to stock one type for all controller thus reducing stock value or saving problems when 

issuing orders to manufacturer. 

Two types of programming menu:  

 

Basic making life easier for those home applications such small swimming pool or water treatment systems; 

 

Expert for professionals giving the opportunity to refine measurements and safety functions. 

A friendly user menu programming it is almost self-explanatory and the elegant wide display allows easy vision. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TECHNICAL CHARACTERISTICS 

Conductivity EC Measuring range and K factor (cel  

constant): 

▪ K 1 ranges up to 20.00 mS and even 100.00 mS (latter with graphite sensor    

electrodes) / Hysteresis / PWM point = 500 μS 

  Resolution 10 μS 

▪ K 5 ranges up to 2.000 μS / Hysteresis / PWM point = 50 μS 

  Resolution 1 μS 

▪ K10 ranges up to 200 μS  / Hysteresis / PWM point = 0,50 μS 

  Resolution 0,1 μS 

▪ K 0.1 ranges up to 100.00 mS / Hysteresis / PWM point = 5000 μS 

  Resolution 100 μS 

(only with graphite sensor electrodes) 

 
Temperature setting: 

automatic with combined conduc PT100 or NTC system  

Temperature probe range: 

– 20 ....100°C 

 ▪   Resolution  0.1% °C        ▪   Accuracy: ± 0.5% °C 

 
Power supply / Power consumption: 

Universal power supply 100

÷

250 Vac / 5W at 240 V AC 

Microprocessor technology: 

SMD components mounting, digital controls keypad 6 key 

Linearity, Stability Reproducibility: 

±

 0.5 % under standard conditions 

Display: 

back-lit custom display (126x64); Display visible area 70x37 mm 

Delay: 

Delay time relay activation, programmable for each set-point (999 sec.) 

Start-up Delay: 

Delay time relay when switching ON, programmable on each set-point 

PARAMETER 

VALUE 

Measure 

Conductivity up tol 20mS  

Type of probe 

PTFE with graphite electrodes, k=1 

Power supply 

100

÷

240 Vac – 50/60 Hz – 6W 

Display 

Graphic 128x64 (visible area 70x37 mm) 

Power supply protection 

The whole power pack (there is no fuse) 

Temperature probe 

PT100, 0,1°C resolution 

External casing 

ABS, IP65 

Inputs 

Level, Proximity 

Relay outputs 

2 relay outputs (set-point 1 and set-point 2) ON-OFF or PWM type

 

Relay outputs features 

maximum 5A on resistive load, maximum 1A on inductive load, 230Vac

 

mA outputs 

1 mA galvanically separated output

 

mA outputs features 

500 ohm maximum load impedence 

Alarm 

Overdosing alarm 

Accessories 

Wall mounting kit 

Measures 

120 x 200 x 90,5 

Weight 

Net weight: 1230 gr  -  Gross weight: 1255 gr 

Front controls 

Polycarbonate adhesive 

Working temperature 

0

÷

50 °C 

Summary of Contents for eSelect-CD1

Page 1: ...EN OPERATING INSTRUCTIONS AND MAINTENANCE FR UNI EN ISO 9001 2008 www etatronds com eSelect CD1...

Page 2: ......

Page 3: ......

Page 4: ...es et lectroniques AEE peuvent contenir des mat riaux nocifs pour l environnement et la sant et doivent donc faire l objet de collecte diff renci e limin s donc aupr s de d charges pr vues cet effet o...

Page 5: ...ng pump relays OUT 1 2 12 DISPLAY DESCRIPTION 13 TERMINAL BOARD ELECTRICAL WIRING 14 OPERATING MODES GLOSSARY 15 PROGRAMMING FUNCTIONS 17 INPUTS OUTPUTS DIAGRAM FUNCTIONS 17 INITIAL DISPLAY 18 Selecti...

Page 6: ...6 ENGLISH SETTINGS EXPERT MENU 29 CONDUCTIVITY SENSORS AND MAINTENANCE 30 CONDUCTIVITY sensors NOTES 30 CONDUCTIVITY sensors maintenance 30 CONDUCTIVITY sensors range 31 TROUBLESHOOTING 31...

Page 7: ...sons or damage systems or materials HIGH VOLTAGE ignoring this sign can result in death or serious injuries to persons NOTE or INFORMATION SOURCE this symbol pointing finger indicates programming step...

Page 8: ...resis PWM point 500 S Resolution 10 S K 5 ranges up to 2 000 S Hysteresis PWM point 50 S Resolution 1 S K10 ranges up to 200 S Hysteresis PWM point 0 50 S Resolution 0 1 S K 0 1 ranges up to 100 00 mS...

Page 9: ...enhanced if it is correctly used and subjected to regular maintenance Warning Qualified and authorized personnel must carry out any intervention or repair for the unit manufacturer declines any respon...

Page 10: ...ns or from the single pole switch contact 2 Take all due safety precautions during unit service gloves goggles overalls etc Assembling and dismantling the equipment ASSEMBLY all ETATRON D S controller...

Page 11: ...CONDUCTIVITY SENSORS with internal temperature sensor due to the various types on the market in case operator wished to use such sensor use PT100 temperature sensors please contact ETATRON D S service...

Page 12: ...s when switching ON OFF try to use a power switch in order to have separate connections via contact relay or relay when this is not possible contact ETATRON technical service EXAMPLE CONNECTION CONTRO...

Page 13: ...CT POWER SUPPLY KEEP IN MIND unit with universal voltage 100 250Vac 10 or 9 24Vdc If the actual voltage is constantly at the limit lower or higher or when spikes are much higher than mentioned range t...

Page 14: ...ime it will appear a list of all the programming keys function Controller covers ranges from costant K 1 up to 20 00 mS and even 100 00 mS latter with graphite sensor electrodes and costant K 5 ranges...

Page 15: ...slight different values between actual measurement in the system and what controller shows USING EXTRA ON OFF RELAY OUTPUT ON OFF Relay probably will connect a dosing pump the free available relays ca...

Page 16: ...of PWM and not 4 sec BENEFITS a simile proportional function more accurate then ON OFF mode using simple equipment such as Constant dosing pumps and peristaltic dispensers DISAVANTAGES not as accurate...

Page 17: ...ed which alarm relay will be ON or OFF mA output It allows to select Conductivity S levels corresponding to min max mA output analogical signal It activates a dosing pump suitable to process a remote...

Page 18: ...ms CONTINUOS MEASUREMENT DISPLAY shows when parameter measurements function status and alarm indications When the controller is turned ON for the first time there will be displayed a list of all the p...

Page 19: ...and results Once selected the programming menu Menu and submenus will change accordingly To help operator to choose the correct menu here following the main difference between the two menus Setpoint...

Page 20: ...se mode works the opposite way Unit is provided with ON OFF mode which will activate or deactivate in case of reverse mode ON output relays such as Constant ON OFF mode dosing pumps or other On Off eq...

Page 21: ...ode dosing pump injects chlorine Oxidant Unit is provided with ON OFF mode which will activate or deactivate in case of reverse mode ON output relays such as Constant ON OFF mode dosing pumps or other...

Page 22: ...sample solution value press OK MENU to confirm Dip EC probe into a buffer solution value which is within the sensor K range wait until measure stabilises To modify the value press OK MENU key then pr...

Page 23: ...o return to CONTINUOS MEASUREMENT display SETPOINT 1 EC CONDUCTIVITY EXPERT MENU Keep in mind that the value displayed is automatically adjusted to the selected conductivity range cell constant K char...

Page 24: ...MIN minimum programmable it sets the minimum time in which PWM is active it overrules the minimum setting programmed In case user has selected Active time min 5 sec this will be minimum time of PWM ch...

Page 25: ...t thus allowing sensors to polarize ensuring correct measurements ONLY WHEN USING DOSING PUMPS Priming function ON will freeze the set point value to allow the dosing pumps to be primed outputs will n...

Page 26: ...it overrules the minimum setting programmed In case user has selected Active time min 5 sec this will be minimum time of PWM check page 15 Timed Pulses function SELECTING MODE ON OFF NEXT PROGRAMMING...

Page 27: ...d is automatically adjusted to the selected conductivity range cell constant K characteristic in previous paragraphs Select the value corresponding to 20 mA point according to Set point 1 settings Pre...

Page 28: ...o confirm selection and to approach next submenu A message will appear before calibration reminding We recommend to make a sample solution which matches the required EC value in the system normally th...

Page 29: ...ess Press ESC to return to MAIN MENU or Press ESC ESC to CONTINUOS MEASUREMENT display FACTORY SETTINGS it restores factory default settings To be used ONLY when parameters or measurements are not res...

Page 30: ...in function of temperature EVERY CONDUCTIVITY SENSOR IS TEMPERATURE DEPENDENT Conductivity varies linearly according to the temperature of the solution This coefficient normalize the measurement of th...

Page 31: ...ls by means of portable instrument or colorimetric kit particularly Temperature reference measure Signal from sensor doesn t change Repeat sensor calibration and if the problem remains change the prob...

Page 32: ...equipment installed relays etc TROUBLESHOOTING HINTS FOR CONDUCTIVITY SENSOR Slow Response typically due to excessive sample line length and low flow thus producing long sample transport lags Resolve...