PROBLEM
SYMPTOM
CHECK THIS
REMEDY
Error message E 01 shows
The cuff is too tight
or too loose.
Refasten the cuff and then
measure again.
E 02 shows
The monitor detected
motion while
measuring.
Movement can affect the
measurement.Relax for a
moment and then measure again.
E 03 shows
The measurement
process does not de
-
tect the pulse signal.
Loosen the clothing on the
wrist and then measure again.
E 04 shows
The treatment of the
measurement failed.
Relax for a moment and then
measure again.
EExx, shows
on the display.
A calibration error
occurred.
Retake the measurement. If
the problem persists, contact
the retailer or our customer
service department for further
assistance. Refer to the
warranty for contact information
and return.
Warning
message
“out ” shows
Out of measurement
range
Relax for a moment. Refasten
the cuff and then measure
again. If the problem persists,
contact your physician.
VI. MAINTENANCE
Dust environment may affect the performance of the unit. Please use the soft cloth to clean
the whole unit before and after use. Don’t use any abrasive or volatile cleaners.
In order to get the best performance, please follow the instructions below.
– Put in a dry (including adaptor) place and avoid the sunshine
– Avoid touching water, clean it with a dry cloth in case.
– Avoid intense shaking and collisions
– Avoid dusty and unstable temperature environment
– Using wet cloths to remove dirt
– Do not attempt to clean the reusable cuff with water and never immerse the cuff in water.
More extensive maintenance or maintenance requiring intervention in the inner parts
of the appliance must be carried out by a professional service!
Failure to comply with the instructions of the manufacturer will lead to expiration
of the right to guarantee repair!