ESWOOD OPERATING MANUAL IW-3N June 2011
6
I
NSTALLATION
4
1.
Remove the lower panel, position the machine and level by screwing the
adjustable feet.
2.
Connect the 2” B.S.P. gravity drain under the machine via a 2” trap to the
drain outlet.
3.
Connect the hot water supply to the ¾” B.S.P. line strainer labelled “HOT”.
The hot water should be at a minimum of 65
°
C and between 100 to 400
kPa. A pressure reducing valve should be fitted for pressure above 400
kPa.
4.
Connect the cold water supply to the ¾” B.S.P. solenoid valve labelled
“COLD”. The cold water supply should between 100 to 400kPa. A pressure
reducing valve should be fitted for pressure above 400 kPa.
If your glassware is sensitive to temperature changes then we suggest that
you do not use the cold water option and cap this solenoid valve
instead.(instruction 4)
Do not make water or plumbing connections that will prevent the
machine from being removed from position, for the purpose of
servicing.
Note that all plumbing work is to be done in accordance with local plumbing
codes.
5.
Ensure that the power switch is “OFF” and connect the three pin plug lead
to a 15 Amp, 240V, 50Hz, 1 phase power outlet.
Note that all electrical work is to be done in accordance with the rating plate
on the lower panel.
6.
The IW-3 comes standard with a detergent injection system that has
already been factory calibrated.
IN RUN POSITION