Bench-top type temperature(humidity) chamber
Installation guide
4
ESPEC CORP.
3. Installation
Installation site check
This section describes installation site and space requirements.
WARNING
Do not install the chamber outside or improper places.
It may cause fire or electrical shock.
Installation requirements
Install in an area that meets the following conditions:
On a flat, level floor strong enough to bear the weight of the chamber
Place subject to minimal mechanical vibrations
Place of minimum airborne contaminants
Place free from moisture
Place free from in flammable materials in the vicinity
Place free from combustible or corrosive gases
Place not exposed to direct sunlight but of good ventilation
Where the ambient temperature is b5 to +35°C, humidity is max. 75%rh
range shown below (+23°C is optimum).
Place where ambient temperature does not change 5°C or more in a few
minutes.
Not directly underneath a fire alarm
Not near a heat source which generates high temperature
Close to power, water and drainage utilities
Altitude not higher than 1000m
Notice
Do not use the chamber under an ambient temperature that is out of the
specified operating range. Doing so may result in malfunction.
The operating condition is b5
C to +35
C.