Installation guide 

Bench-top type temperature(humidity) chamber 

 

ESPEC  CORP. 

15 

 

 

Power supply voltage (V AC)  Max. current (A) 

Min. power 

cable size 

(mm

2

Fuse capacity 

(A) 

SH/SU- 

  642,662 

100V 50Hz/60Hz 1Ø 2W 

17.5 (18.5) 

20 

200V 50Hz/60Hz 1Ø 2W 

14.5 (15.0) 

15 

220V 50Hz/60Hz 1Ø 2W 

14.0 (14.5) 

1.5 

15 

230V 50Hz 1Ø 2W 

14.0 (14.5) 

1.5 

15 

SH/SU- 

  242-5 

100V 50Hz/60Hz 1Ø 2W 

16.7 (17.7) 

20 

200V 50Hz/60Hz 1Ø2W 

10.6 (11.1) 

15 

220V 50Hz/60Hz 1Ø2W 

9.3 (9.8) 

1.5 

15 

230V 50Hz 1Ø2W 

9.2 (9.7) 

1.5 

15 

 

* Figures in parentheses are for when a viewing window is installed. 
* Option:115V,200V,220V,230V AC. 

 

 

WARNING

 

 

When using a leakage beaker as the primary power switch, use one that 
supports high harmonics. 

 

Malfunctions can occur with switches that do not support high-harmonics. 

 
Primary power supply connections 

 
The chamber is connected to the primary power supply via the power cable. It is also 
grounded  to  prevent  noise  from  infiltrating  electronic  circuits  and  to  protect  users 
against electric shock which can result from leakage current. 

 

WARNING

 

 

EXPLOSION! DO NOT ground the equipment on gas pipes. 

Grounding of the sort could result in explosion. 

 

 

DO NOT share ground with equipment not fitted with a leakage breaker. 

The  chamber's  leakage  breaker  will  not  trip  in  the  event  of  overcurrent, 
short-circuits or surges. 

 

DO NOT ground between equipment. 

 

The  chamber's  leakage  breaker  will  not  trip  in  the  event  of  overcurrent, 
short-circuits or surges. 

 

Ground the equipment WITHOUT FAIL 

 

Unless the equipment is grounded, the leakage breaker will not trip in the event 
current leaks, possibly resulting in electric shock. Furthermore, this will leave the 
equipment  unprotected  against  power  supply  noise.  Sometimes,  the  temperature 
(humidity) in the chamber can be destabilized. 

 

 

 

Summary of Contents for SH-222

Page 1: ...ORP Installation guide Bench top type temperature and humidity chamber SH SU 40093010010A0 May 10 2019 This guide describes how to bring in and install the chamber Before starting these works read it...

Page 2: ...ure to observe handling instructions contained herein Do not perform any operation or handle the chamber in any way or form that is not described in this guide or which is herein specifically prohibit...

Page 3: ...EC CORP 1 Contents 1 Safety precautions 2 2 Manuals 3 3 Installation 4 4 How to remove moisture absorbents export specifications only 9 5 Checking accessories and spare parts 10 6 Drainage work 12 7 P...

Page 4: ...ath or serious injury to the user if the chamber is handled improperly CAUTION Means that dangerous consequences may arise with the risk of minor injury or light wound to the user if the chamber is ha...

Page 5: ...l Organization for Standardization ISO The certification criterion for green product labeling is as follows Products which are intended for the domestic market and satisfy the following requirements A...

Page 6: ...free from moisture Place free from in flammable materials in the vicinity Place free from combustible or corrosive gases Place not exposed to direct sunlight but of good ventilation Where the ambient...

Page 7: ...g an appropriate temperature in the installation site reduces the load on the refrigerators and heaters imposed by the ambient temperature Notes for installation List of Center of Gravity Positions WA...

Page 8: ...r door Dimensions in brackets include protrusions instrumentation frame condenser fan motor and cover for compressor terminals Protrusions instrumentation frame condenser fan motor and cover for compr...

Page 9: ...222 242 440 690 747 440 SU 222 242 440 625 747 440 SH 262 440 690 834 440 SU 262 440 625 834 440 SH 642 662 540 730 943 540 SU 642 662 540 665 943 540 SH 242 5 440 690 834 440 SU 242 5 440 625 834 440...

Page 10: ...evel floor An unlevel chamber can possibly cause water to backflow without smooth drainage from the dew tray mistakenly trigger alarms and destabilize humidity control Leveling feet are found under ea...

Page 11: ...sture absorbents export specifications only The chamber door is fitted with a retainer before shipping from the factory so as to prevent the door from opening during transport After the chamber has be...

Page 12: ...wick Cartridge fuse Class CC fuse Breaker handle cover Stylus pen Operation manual Warranty card No Accessory Part Use Q ty Check if included Shelf Holds specimens inside the chamber 1 Cable port rubb...

Page 13: ...ion circuits and parts SH SU 222 242 100 220 230V AC B type 250V 7A 115V AC NEC spec UL compliant 250V 0 75A 0 5A 0 06A 1 1 each SH SU 242 5 B type 250V 7A 6A 1 each SH SU 262 642 662 Except NEC spec...

Page 14: ...iner To prevent backflow into the chamber do not share the drainage line with other equipment Also leave the water tank drain hose and drain pipe ends open to the atmosphere Raising the drain pipe abo...

Page 15: ...n port A socket for draining when cleaning the water supply pump filter SH only Humidifying tray drain port A socket for draining when performing temperature operation or cleaning the humidifying tray...

Page 16: ...ly 100V AC 1 2W 50 60Hz 115V AC 1 2W 60Hz 200V AC 1 2W 50 60Hz 220V AC 1 2W 50 60Hz 230V AC 1 2W 50Hz Power supply voltage fluctuation 90V to 110V 103V to 126V 190V to 220V 198V to 242V 219V to 241V 1...

Page 17: ...tions The chamber is connected to the primary power supply via the power cable It is also grounded to prevent noise from infiltrating electronic circuits and to protect users against electric shock wh...

Page 18: ...h ground The chamber is grounded through the power cable when plugged into the electrical outlet so separate grounding work is not needed 220 and 230V AC supplies have a cable with the below plug Prep...

Page 19: ...200V AC Black White Green 220V 230V AC Black Black White Green Yellow For SH SU 222 242 115V AC Power cable color Power supply voltage L phase N phase GND 115V AC Black White Green Yellow To primary p...

Page 20: ...the chamber 2 Press the instrumentation power switch Power OFF Procedure 1 Press the instrumentation power switch 2 Turn the circuit breaker OFF Note Do not repeatedly turn the circuit breaker ON and...

Page 21: ...onductivity between 0 1 to 10 S cm Using water with lots of impurities such as tap water will shorten life time of the humidifying heater and wet bulb wick Do not use ultra pure water Ultra pure water...

Page 22: ...ty from attaining or keeping the set targets Do not duplicate setting water tank to water circuit compartment for many times It may create a loss of pressure balance As the result siphon effect can oc...

Page 23: ...key for Hum Control The dot on the key lights up yellow To set the humidity press the target setting box for Set Hum The temperature input range appears in the pop up window Input 23 C for the temper...

Page 24: ...o the humidifying tray will be completed within about 5 minutes 8 Make sure it is over 5 minutes After the water supply is completed finish the constant value operation Press the Stop button below Sto...

Page 25: ...ve or below the WATER LEVEL mark water is overflowing or the humidifying heater is sticking out of the water in the humidifying tray the following may have happened The water tank is not properly conn...

Page 26: ...ulator is above the WATER LEVEL mark 3 If water in either the humidifying tray or wick pan is off the WATER LEVEL mark Loosen the nut and turn the water level sensor clockwise seen from above to lower...

Page 27: ...MEMO...

Page 28: ...Edited and Published by ESPEC CORP 3 5 6 Tenjinbashi Kita ku Osaka 530 8550 Japan Tel 81 6 6358 4741 Fax 81 6 6358 5500 Printed in Japan...

Reviews:

Related manuals for SH-222