SETUP CAUTIONS
Duri ng the setup phase it is conven ient to i ssue a STATUS command so
the process of bi n setup is visible on the screen .
Remember that
when the setup i s complete, STATUS must be i ssued agai n to togg le the
i nstrument back to the chosen measurement mode. Otherwise, the S I NGLE
measurement
commands w i l l
not
result
i n
the
transmiss ion
of
measurement data to the remote devi ce.
To al low fastest setup times and a lso to ensure that the remote i nput
w i l l n ot be d i sturbed by keyboard or measurement i nput between
successive l i nes of setup i nformation, the f i rst character output
( i nstructi on i n the program) should be :
REMOTE[
whi ch temporari ly discon nects the keyboard and external START switch
( i. e. do not use SI NGLE or CONT I N UOUS). This entry must be on a l i ne
by itself fol lowed by a carriage return before additional i nput w i l l
be a c cepted by the
2 15 0/2 160.
The
"
J
"
character is issued at the end of the setup phase to restore
the i nstrument to its measurement loop and make the keyboard active
again.
" J "
automat i ca l ly transm its error i nformation. The
must
lock the
out if that is des ired.
Th is may
be
done with the
LOCK
command.
After the
measurements have been made, the i nstrument can be UNLOCKed.
elsli
A
-
35
2 15 0/2 160
9 /85
desi red
Summary of Contents for VideoBridge 2150
Page 24: ...BLANK 1 6 e l s l i 2 1 50 2 1 60 5 8 5...
Page 230: ...R54 R 1 9 R23 6 TP8 o O NC N N N N a a a a a F i gure 5 5...
Page 252: ...A t S E C T I ON A A 4 5 0 0 m m 5 rY1 m A...
Page 307: ...Figure 8 1 General Handler Interface Circuit Assembly P N 47895 B 2 e s i 2150 2160 5 85...
Page 330: ...F i gure 8...