29
Class II Biological Safety Cabinets
4.
Remove temperature sensor from the bracket.
Figure 5.9. Remove Temperature sensor
5.
Remove the doughnut sensor housing from the bracket.
Figure 5.10. Remove doughnut sensor housing
6.
Install the new sensor (it comes with the doughnut sensor housing) into the bracket and connect the
sensor cable connector. Ensure the airflow direction described in the sensor housing is complied with
during installation.
Figure 5.11. Airflow sensor direction