Problem 7: Light always OFF
Cause
Corrective Action
Sash in SASH
ALARM state
•
Is the Light LED on the front panel ON?
•
If the light does not turn on, press the Light button and make sure the sash is in the
Normal operating position (Ready Position Normal Operating height).
Note:
The sash must be at the Normal Operating position
.
•
Is the cabinet in Warm Up period?
Note:
The fluorescent light will not light if the cabinet is in Warm Up period unless
the cabinet is in Quick Start Mode.
•
Move the sash to Ready position (normal operating height).
•
Switch on the light by pressing LIGHT button on the membrane.
•
If the LED is ON the keypad and main board are functioning correctly.
•
If fluorescent light does not light with the window in the proper position Proceed to
the next step
•
If the LIGHT LED still does not light, skip to
Ready magnetic switch Issues
in this
table.
Burned out
fluorescent tube
•
Replace the faulty fluorescent tube.
•
Fluorescent tubes are located inside the blue panel.
•
The white lamp holder can be pulled off from the tubes.
Defective
Fluorescent ballast
or Relay Board
issues
•
Check incoming voltage to the Relay board.
•
Locate LS7 circuit and measure AC voltage between BLUE (NEUTRAL) terminal block
and the end terminal (with Red cable) on the three terminal strip (J13 on relay
board).
•
If there is voltage to the end terminal (Terminal with cable), check the central
terminal (Normally Open Circuit) to the BLUE (NEUTRAL) terminal block.
•
No voltage at the central terminal indicates a blown fuse or the circuit is not
energized.
•
Check Fuse F7 on relay board. See Figure 7-1 below.
•
If fuse F7 is blown, as temporary solution, use F5 (spare) to replace F7.
•
If fuse F7 is good, check the voltage between the BLUE (NEUTRAL) terminal block
and the end terminal on the LS7 circuit (J13 on relay board).
•
Voltage at the end terminal with no cable indicates the F7 fuse is good and the
circuit is not energized.
•
Make sure the LIGHT LED is on.
•
If the ballast has power and the bulbs are new the problem is either connections or
the ballast.
•
Check Connector C connections. If there are no lose terminals, replace the ballast.
Ready magnetic
switch Issues
•
With the LIGHT LED ON and the sash at the Normal operating position (Ready)
•
Does the Display indicate “
SASH: OK
” or “
SASH: NO!
”
•
“
SASH: NO!
” indicates the sash is not at the proper position and the LIGHT LED
should not be on. This indicates a main board problem or a magnetic switch
detection problem.
Figure 7-1
F7
LS7
Summary of Contents for Class II
Page 26: ...16...
Page 64: ...54 8 2 AC2 _S_ Engineering Drawing...
Page 66: ...56 8 4 AC2 _D_ Engineering Drawing...
Page 67: ...57 Class II Biological Safety Cabinets...
Page 68: ...58 8 5 AC2 _G_ Engineering Drawing...
Page 70: ...60 8 7 AC2 2N7 Engineering Drawing A C 2 2 N 7 E n g i n e e r i n g D r a w i...
Page 71: ...61 Class II Biological Safety Cabinets...
Page 73: ......