voltmeter. If voltage is not /-2% of the socket voltage, replace the 
power cord.  

Check the fuse of the power cord.  

 

 

 

CO2 level is below or above CO2 set-point  

Possible Cause  

 

 

 

Corrective Actions  

Unit is not connected to the pressure 
regulated CO₂ source.  

Check gas hose connection at the rear of unit 
to the CO2 source as defined in the 
installation section of the manual.  

 

CO2 inlet pressure not regulated correctly  

The CO2 inlet gas pressure should be 
regulated to approximate 15 psig.  

 

CO2 gas tank is empty  

 Replace CO2 gas tank.  

 

In-line filter is dirty / saturated 

Replace in-line filter.  

 

Gas hose is leaking  

If hose is leaking, check the pressure regulator 
to ensure CO2 inlet pressure is not more than 
15 psig. Check for leakage at hose connection.  

Gas hose is dirty or obstructed  

Check physically for dirt accumulation or 
obstruction, clean or replace it.  

 

CO2 set-point was recently lowered or raised 
Unit is still not stabilized after the set-point 
was changed  

Check CO₂ set-point. Allow 2 hours for the 
unit to stabilize after adjustment of new set-
point.  

If the CO2 set point is lowered, the outer and 
inner doors must be opened.  

 