HM 

 
HEAT LED 

 

 

HEAT LED will light up whenever a heating process is activated  

 

INJECT LED 

 

 

INJECT LED will light up whenever a gas injection process is activated  

 

oC/%RH Display 

 

 

When oC displayed, the display shows the temperature in the chamber  

 

When %RH displayed, the display shows the relative humidity in the chamber 
(optional)  

 

%CO2/%O2 Display 

 

 

When %CO2 displayed, the display shows CO2 concentration level in the 
chamber  

 

When %O2 displayed, the display shows O2 concentration level in the chamber 
(suppressed O2 model)  

 

 

 

Problems / Corrective Actions  

 

Unit will not turn on 

 

 

Possible Cause  

 

Corrective Actions  

Power failure  

 Is the cord set connected to the power inlet at rear of cabinet as defined 
in the installation section of the manual?  

 Is there power at the electrical outlet which the unit is plugged into?  

 Use voltmeter or test pen to check power on the electrical socket.  

 Is the unit’s power switch turned on?  

Check whether power cord has been connected properly into the 
wall/building socket and also into the incubator.  

Check whether the power cord is giving power, measure the AC voltage 
between the live and the neutral terminal of the cord by using a 