AVT-1
®
User Manual Rev. 1.8
13
It is the fully laden float that should be level, thus the adjustment is dependent on the
microscope weight.
The float height is adjusted by loosening 4 securing hexagonal head bolts and turning
the handle.
Figure 7.6
Adjusting the float height
After adjustment, tighten the four securing bolts again. If this is done improperly, the
table will go out of adjustment while in use.
Any additional equipment may now be installed.
Make sure that all cables that are connected to the microscope hang freely between the
attachment point on the microscope and the other side. This can be achieved by strip-
ping the cable to the points and making sure that the cable forms an arch between the
points. If the cable is tight or touches area between the anchor points, vibration may be
transferred via the cable.
After successful installation and device testing, complete user training and instructions.
8 Cleaning instructions
Always perform cleaning procedures locally; for more guidance consult ei-
ther your manufacturer or the distributor.
Switch off the devices on the table and remove the mains leads. Clean the surfaces of
AVT-1
®
with a 70% alcohol solution, using a lint-free cloth.
A list with approved cleaning liquids can be supplied on request.