T E C H N I C A L S P E C I F I C A T I O N S
MyLab
- G E T T I N G S T A R T E D
9 - 6
Weight
< 65 kg (basic configuration without peripheral units)
IP Grade
IP (X)0, this means
MyLab
models are not watertight.
Power supply
Voltage operative range:
• 100
120 V
• 200
240 V
Working frequency range: 50
60 Hz ±10%
Power consumption:
220 VA (
MyLab
only)
Power consumption:
600 VA (
MyLab
+ peripherals)
Available power on peripherals: up to 250 VA
Fuses: T 5Ah, 250 V (220V)/ T 10Ah, 250 V (110V), 5x
20mm
Batteries
Batteries for:
• stand-by
suspension
• fast movement without switching off
Battery charger inside
Battery charging cycle: around three and a half hours (3 h 30
min)
Battery life: 3 years
Summary of Contents for MyLabX6
Page 1: ...Rev 08 February 2020 MyLabX6 MyLabX7 GETTING STARTED 350031500...
Page 2: ...MyLab G E T T I N G S T A R T E D ii...
Page 5: ...MyLab G E T T I N G S T A R T E D v GETTING STARTED EC Declaration of Conformity...
Page 6: ...MyLab G E T T I N G S T A R T E D vi...
Page 7: ...MyLab G E T T I N G S T A R T E D vii GETTING STARTED RED Declaration of Conformity...
Page 8: ...MyLab G E T T I N G S T A R T E D viii...
Page 9: ...MyLab G E T T I N G S T A R T E D ix GETTING STARTED...
Page 10: ...MyLab G E T T I N G S T A R T E D x...
Page 16: ...I N D E X MyLab G E T T I N G S T A R T E D xvi...
Page 28: ...I N T R O D U C T I O N MyLab G E T T I N G S T A R T E D 1 12...
Page 49: ...S Y S T E M O V E R V I E W MyLab G E T T I N G S T A R T E D 3 13 GETTING STARTED...
Page 66: ...P R E P A R I N G T H E S Y S T E M MyLab G E T T I N G S T A R T E D 4 12...