50
By the numerical keyboard, set the desired initial scale value (For
Example 3 it is equal to 4.000 mA).
NOTES
:
1) The value assigned to "H.RANGE mA" must be higher than the
value assigned to "L.RANGE mA" parameter.
2) The value assigned to "H.RANGE EU" parameter must be
different from the value assigned to "L.RANGE EU".
When conditions 1 or 2 are not satisfied, the instrument will reject
the new value showing the "0" and it will maintain the cursor on the
rejected parameter.
3) The value assigned to "H.RANGE EU" parameter can be lower
than the value assigned to "L.RANGE EU" (reverse read-out).
Push the FUNC push-button; the cursor will go automatically on the
lower line.
By the numerical keyboard, set the desired read-out assigned to
the initial scale value (For Example 3, it is equal to 5.00 m
3
/h.)
Depress the FUNC push-button.
NOTES
:
1) The input range may contain negative values.
2) The displayed range will only be represented with positive
numbers.
The display will show the following:
By the numerical keyboard, set the desired final scale value (For
Example 3 it is equal to 20.000 mA).
Push the FUNC push-button.
By the numerical keyboard, set the desired read-out assigned to
the final scale value (For Example 3 it is equal to 200.00 (m
2
/h).
Push the FUNC push-button.
NOTES
:
1) For the read-out in eng. units, the number of decimal figures will
be equal to the decimal figure programmed in "L.RNG EU".
2) For the read-out in eng. units, the number of decimal figures is
free with respect of the read-out limits (20000 counts).
The instrument will start to measure and, for Example 3, if a 7 mA
signal is applied to the input terminal the display will show:
NOTES
:
1) The instrument will automatically select the input range contain-
ing the two programmed values ("L.RANGE mA" and "H.RANGE
mA") with the maximum resolution.
L.RNG XXXXXX mA
L.RNG XXXXXX EU
H.RNG XXXXXX mA
H.RNG XXXXXX EU
MEAS RNG( )mA
86.68 EU
M20-e-3.p65
7/9/01, 2:40 PM
50