32
By the numerical keyboard, set the desired initial scale value (from
Example 2, this value is equal to 4.000 mA).
NOTES
:
1) The value assigned to "H.RANGE mA" must be higher than the
value assigned to "L.RANGE mA" parameter.
2) The value assigned to "H.RANGE EU" parameter must be
different from the value assigned to "L.RANGE EU".
When conditions 1 or 2 are not satisfied, the instrument will reject
the new value showing the "0" and it will maintain the cursor on the
rejected parameter.
3) The value assigned to "H.RANGE EU" parameter can be lower
than the value assigned to "L.RANGE EU" (reverse read-out).
Push the FUNC push-button; the cursor will go automatically on the
lower line.
By the numerical keyboard, set the desired read-out assigned to the
initial scale value (for Example 2 this value is equal to 0.0 l/h).
Depress the FUNC push-button.
NOTES
:
1) The displayed range will only be represented with positive
numbers.
2) The instrument selects the 20 mA input range.
3) For the read-out in eng. units, the number of decimal figures is free
with respect of the read-out limits (20000 counts).
The display will show the following:
By the numerical keyboard, set the desired final scale value (For
Example 2, this value is equal to 20.000 mA).
Push the FUNC push-button; the cursor will go automatically on the
lower line.
By the numerical keyboard, set the desired read-out assigned to
the final scale value (for Example 2, this value is equal to 500.0l/h).
NOTE: For
the read-out in eng. units, the number of decimal
figures will be equal to the decimal figure assigned to
"L.RANGE EU".
Push the FUNC push-button.
The instrument will start to measure .
For example 2, if a signal equal to 9.718 mA is applied to the input
terminal the display will show:
NOTES
:
1) During a TX measurement it is possible to enable/disable, by
the REV/MOD push-button, the scaling function and to display
the measured value in mA.
2) During TX measurement, the instrument is able to measure and
display up to 5 % more than the upper programmed value or 5%
less than the lower programmed value.
H.RNG XXXXXX mA
H.RNG XXXXXX EU
MEAS TX(
)
298.9 EU
M20-e-3.p65
7/9/01, 2:40 PM
32