15 - TROUBLESHOOTING
Problem
Possible Issue and Solution
No air to workstation
Air compressor is off
Air supply valve is off
Air regulator not adjusted properly. Adjust to 80-85 PSI
Master switch is off or not suppl;ying air to master regulator
No bottle water supply
Master switch is off.
Bottle is empty
Bottle/ Line switch is on line (if applicable)
Bottle Pressure switch is off (if applicable
No or low drive air for the handpieces check air pressure supplied to workstation at 80+ PSI.
Adjusting screws on bottom of power channel not adjusted properly
No water for handpieces
Check water supply with syringe, if no go to 2
Wet/Dry switch on face plate or foot pedal not switched on
Flow control for handpieces not adjusted
No air / water at syringe
Check air supply, bottle water supply
Workstation does not turn on
Verify that the workstation is properly connected to a working air source
No line water supply (optional)
Dental water supply switch (on Workstation) or water valve shut off
at regluator
Water regulator not adjusted
Master switch is off or not supplying air to master regulators.
For Support, call Ergonomic Products directly at 1-866-374-6487