Appendix
86/96
16.1.2
Table 202: Guidelines and manufacturer's
declaration – Electromagnetic interference
The Tanning system has been designed for use in the electromagnetic envi-
ronment specified below. The device operator must ensure that it is used in
such an environment.
Interference
immunity tests
IEC 60601 -
Test level
Defined
level
Electromagnetic
environment –
Guidelines
Discharging static electricity
(ESD) IEC 61000-4-2
±8 kV contact discharge
±2 kV air discharge
±4 kV air discharge
±8 kV air discharge
±15 kV air discharge
±8 kV contact discharge
±2 kV air discharge
±4 kV air discharge
±8 kV air discharge
±15 kV air discharge
–
Fast transient electrical noise /
bursts IEC 61000-4-4
±2 kV for mains supplies
±2 kV for mains supplies
–
Surges
61000-4-5
±1 kV series mode voltage
±2 kV common mode voltage
±1 kV series mode voltage
±2 kV common mode voltage
–
Magnetic field radiated by
supply frequency (50/60 Hz)
IEC 61000-4-8
30 A/m
30 A/m
–
Voltage drops,
Short interruptions
and supply voltage
IEC 61000-4-11
<5% U
T
(>95% drop in U
T
)
for 5 s
Only this test is performed,
because the device used 2 or 3
phases and has more than16A /
phase.
<5% U
T
(>95% drop in U
T
)
for 5 s
–
Note: U
T
is the mains alternating voltage used for the test levels.