Cleaning and maintenance
44
5.8
Instructions for replacing the lamps
DANGER!
Danger of being burnt by an over-powered lamp!
If the old chip card is left in the device, the new lamp set will
be run at too high power and the user may suffer heavy
burns!
–
The enclosed chip card must always be used when
changing the UV low pressure lamps!
The lamp's UV effect and therefore the tanning effect decrease over time. In
electronically controlled low pressure lamps, this loss of effect can be com-
pensated. The chip card is used to control the changing periods. You must
adhere to the replacement intervals specified in the maintenance plan.
UV filters are used in order to control the UV radiation, e.g. the filter panels in
the facial tanner. However, the filter effect decreases over the course of time.
You must adhere to the replacement intervals specified in the maintenance
plan.
5.8.1
Changing the chip card
(AFFINITY 660-S DYNAMIC POWER /
AFFINITY 880-S DYNAMIC POWER)
The chip card belongs to the long UV low pressure lamps fitted in the canopy,
the side panels and the device's base.
The chip card is located in the 'Part support control'.
–
Opening the bed base: see page 43.
–
Remove the chip card belonging to the old lamp set.
–
Insert the new chip card with the arrow pointing downwards.
–
Close the bed base. You should hear the bed base click into place.
Summary of Contents for AFFINITY 500-S SUPER POWER
Page 2: ......
Page 30: ...Bedienung 28 4 5 bersicht Bedienung...
Page 82: ...Anhang 80 5 Aufkleber Wartungshinweis 800701...
Page 88: ...Index 86...
Page 116: ...Operation 28 4 5 Operating overview...
Page 168: ...Appendix 80 5 Maintenance instruction sticker 800701...
Page 174: ...Index 86...
Page 175: ......