background image

 

 

 

 

 

5)  Hang the electrical box-system on the inner side of the handle. 

6)  Please press the red emergency button on the electrical box before connecting motor and box. The 

emergency button disconnects all electrical contact. You then can connect the motor’s connection cable to 
the 2 pin plug on the underside of the electrical box. The control device is connected to the plug on the 
back of the electrical box using the 3 pin plug. Release the emergency button. Leg spreading (electric) – 
connect the 8 pin plug to the corresponding socket underneath the electrical box. 

8)  Ensure that the emergency button is in the extended position. If not, give the button a quarter turn 

clockwise until it reaches the extended position. 

1.4 Purpose 

The Pallas 2000  +  Pallas  5000 is designed  to  lift persons  in nursing care  and  with at  least one instructed
carer  or  supervisor.  The  load  capacity  is  150  kg. 

Never  ever 

use  the  hoist/lift  for  other  purposes  than 

originally intended. 

 

 

1.5 Use 

The hoist/lift should mainly be used in the following situations: 

1.  Lifting the user from a bed to a wheel chair/chair situated immediately beside the bed. 

2.  Lifting from a wheel chair to toilet/toilet seat situated in the immediate proximity. 

3.  Lifting from a wheel chair to a chair situated in the immediate proximity. 

Use only the slings exclusively recommended. 

1.6 General remarks concerning use 

The hoist/lift is easy to operate. However, before use, please read the following instructions carefully: 

1.  The legs of the chassis can be spread with the foot pedal or alternatively with the control device for 

electrical leg spreading models. 

2.  The control device enables easy handling. Just press gently on the button “up” or “down”. Electrical 

leg spreading model – Press gently on the corresponding buttons. 

3.  Always ensure that the user feels safe and is not frightened. Wherever possible ensure close contact 

with the user. 

4.  Pressing the 

red emergency button 

immediately stops the hoist’s/lift’s function. By turning a 1/4-turn 

clockwise the button extends and the hoist/lift is again ready for use. 

 

 

 

5. The hoist/lift does not excert downward pressure other than the  weight of the user and  arm combined. 

This safety feature fitted to the spindle is called “spline” and reduces the danger of accidental squeezing. 
In very extreme circumstances the spline system may have been activated and can result  

in the lifting arm falling into place. Whilst this does not compromise safety it can be disturbing if a client is 
being  lifted  and  then  suddenly  falls  into  place.  We  recommend  that  the  lift/hoist  is  operated  for  a  short 
distance before using with a client, to ensure that the spline system is in place. 

6.  The Pallas 2000 has an 

electronic button 

placed underneath the electrical box making an emergency 

lowering possible, in case the control device is defect. The Pallas 5000 has clearly marked emergency 
points (lowering/raising) on the electrical box. 

7.  Please charge batteries regularly - we recommend every night - and ensure that the emergency button 

is  in  the  extended  position.  Do  not  wait  between  two  charging  processes  until  the  light  diode  (lamp) 
shows 50%  discharge. The batteries do not have any “memory effect”. Their lifetime can be critically 
reduced if allowed to go flat before recharging. 

 

 

Summary of Contents for Pallas 2000

Page 1: ...USER GUIDE USER GUIDE USER GUIDE USER GUIDE Pallas 2000 5000 Pallas 2000 5000 Pallas 2000 5000 Pallas 2000 5000 Version 5 22 november 2007...

Page 2: ...ifting with the universal sling 10 3 3 Pillow sling 10 3 4 Use as a walking aid 11 4 SERVICE 11 5 5 1 5 1 5 2 5 3 5 4 5 5 TECHNICAL FUNCTIONS Charging the batteries 1Charging the batteries speed Elect...

Page 3: ...d relative humidity 40 90 The lift load must not exceed 150 kg Before use ensure that the hoist lift has been assembled correctly and according to these instructions Use only slings mentioned in this...

Page 4: ...ctuator 1 electrical box 1 control device 1 charger 1 3 Assembly For assembling the hoist lift you need two 17 mm spanners included Please follow the advice in the order mentioned 1 Place the mast wit...

Page 5: ...or alternatively with the control device for electrical leg spreading models 2 The control device enables easy handling Just press gently on the button up or down Electrical leg spreading model Press...

Page 6: ...er than necessary The Pallas 2000 Pallas 5000 are not means of transport It is the responsibility of the carer to ensure that this product is suitable for the individual client The utmost care must be...

Page 7: ...TL Small Medium Large L Large Head Neck support TL One size 000 03514 000 03513 000 03511 000 03512 007 03530 Universal sling TL Small 001 03554 Medium 001 03553 Large 001 03551 X Large Universal slin...

Page 8: ...move the hoist lift to the user Lift the user s feet onto the foot plate bring the hoist lift forwards until the knee support is very close to the user s knees The height of the knee support can be a...

Page 9: ...as possible and look forwards Ask the user to press down firmly onto the foot plate and to stretch both knee and hip joints The carer may help by placing a hand on the knee and press the knee lightly...

Page 10: ...outer side of the thighs Then place them under the thighs and cross the sling straps in front The user may be raised in a more or less reclining position e g a relatively short back strap and a relati...

Page 11: ...ton is in the extended position Do not wait between two charging processes until the light diode lamp shows 50 discharge The batteries do not have any memory effect Their lifetime can be critically re...

Page 12: ...y be cleaned with a damp cloth in order to avoid the electronic components being damaged The rest of the Pallas 2000 Pallas 5000 may be cleaned with detergent or disinfectant 6 GUARANTEE The Pallas 20...

Page 13: ...of the knee support and the foot plate 23 55 cm Diameter of wheels 75 1 00 mm Clearance under chassis 5 cm Chassis min Internal width 45 cm Chassis max internal width 72 5 cm Hoist reach max 80 cm He...

Page 14: ...harger with spiral cable 30V 0 5A 253 00100 Charger regulated 24 VDC 1 0 Ah LG 624 253 00201 Charger Jumbo 750 63021 Foot pedal with mounting parts 018 03192 Tente wheel with brake 100 mm 2 pcs 018 03...

Page 15: ...018 00603 Safety rubber set s crew without plastic cap 018 00500 Foam for arm long old model 018 00501 Foam for h andle long 018 00502 Foam for arm short model 0 18 00503 Foam for arm divided in two...

Page 16: ...Defect print card Replace control device Replace print card 4 Periodical fault The lifter travels either up or down not both or spreads legs either in or out not both Power shortage or fault in the e...

Page 17: ...rol clockwise will increase the speed Charging Beeps when the charger is connected The diode which is placed adjacent to the emergency stop will blink green during the charging process The same diode...

Reviews: