18
5 CHECK AND TROUBLESHOOTING
5.1 Annual check
Pallas 175 must always be maintained by qualified and authorised personnel who are been
approved by Winncare Nordic.
Maintenance involves
: Replacing batteries, charger, lifting straps and other components.
The manufacturer Winncare Nordic stipulates that this hoist MUST be checked by a qualified
and authorised person at least once a year. This annual check is also recommended in the
relevant European Standard (EN ISO 10535) for patient lifters/hoists. More frequent checks
may be required by national laws.
It is the onus of the buyer to ensure that this annual check is carried out.
If it is necessary to replace any parts due to wear or damage these must be
obtained/purchased from Winncare Nordic.
5.2 Troubleshooting
Fault:
The lifter does not
react when the hand
control buttons
UP/DOWN are used.
Possible cause:
The lifter is not on or emergency stop is
activated.
Solution:
Deactivate emergency stop. (see section 2.4)
Possible cause:
Defect hand control.
Solution:
Try using another hand control. Alternative is
to get a new hand control from the dealer.
Possible cause:
Flat battery.
Solution:
Charge battery. (see section 2.5)
Fault:
The lifting arm can
travel up or down not
both.
Possible cause:
Fault in the electrical control box or control
device.
Solution:
Replace the control box or control device.
Summary of Contents for Pallas/175