17
Luna may be cleaned using a slightly damp cloth with a small quantity standard household
cleaning agent. Chemicals and other strong/abrasive agents must not be used.
Luna cannot under any circumstance be autoclaved.
Luna must be stored in a dry environment with relative humidity under 70%.
If the Luna is not to be used over a longer period of time it is best to fully charge the unit
and activate the emergency button to save battery before putting in storage.
Luna must always be maintained by qualified and authorised personnel who have been
approved by Ergolet for replacing batteries, lifting straps and other components
The manufacturer Ergolet stipulates that this track lift MUST be checked by a qualified and
authorised person at least once a year. This annual check is also recommended in the
relevant European Standard (EN ISO 10535) for patient lifters/hoists including track lifters.
Ergolet’s service personnel will fill out the logbook after each annual check. If it is
necessary to replace any parts due to wear or damage these must be obtained/purchased
from Ergolet.