Resetting the “Next Safety Check” Date on a VIO 300D ESU

 

 

 

 

2225 Northwest Parkway • Marietta, GA 30067 • Tel: (800) 778-3723 • Fax: (770) 955-2577  

  

www.erbe-usa.com

 

 

Rev. Sept. 2016

 

 

 

Figure 3 

9.

 

Review the current “Date and Time” options; make any corrections at this time by simply select/highlight 
the setting that requires adjusting. Using the buttons on the gray frame, press the “Up” and/or “Down” 
buttons to adjust the settings. 

10.

 

Once date and time are verified and adjusted as required, press the “Confirm/Enter” button (

See Figure 

3

) to exit the Date or Time selections. 

Note: Step # 10 is accomplished when the word “

More

” reappears on the glass display. 

11.

 

Press “More” several times until “Next Safety Check” is visible.  

12.

 

Select “Next Safety Check”. 

13.

 

Using the buttons on the gray frame to the right of the glass display, press the “Up” button one time. 

(See Figure 4)

 

Observe that the “Year” advances one year from the current year.

 

 
 

 

Figure 4 

14.

 

Press the Confirm/Enter button to save the change. 

(See Figure 3 above)

   

15.

 

Turn “Off” the VIO 300D, then “On” again. Ensure that the “Next Safety Check” green message does not 
appear. 

 

FOLLOW THESE STEPS WHEN THE REMINDER MESSAGE “NOT” PRESENT 

 

Note: All button selections will be made with the buttons on the glass display of the unit unless 
otherwise stated. 