Resetting the “Next Safety Check” Date on a VIO 300D ESU
2225 Northwest Parkway • Marietta, GA 30067 • Tel: (800) 778-3723 • Fax: (770) 955-2577
Rev. Sept. 2016
FOLLOW THESE STEPS WHEN THE REMINDER MESSAGE IS PRESENT
Note: All button selections will be made with the buttons on the glass display of the unit unless
otherwise stated.
1.
Turn “On” the VIO 300D. Allow the “green” message box to disappear.
2.
Select “Return”.
3.
Select “Other Functions”.
4.
Select “Setup”.
5.
Press “More” until the “Service Program” option appears.
6.
Select “Service Program”.
Note: The Password Screen will appear. The password shortcut is the dash symbol, “-“.
7.
Using the buttons on the gray frame to the right of the glass display, press the “Down” button one time.
(See Figure 1)
Figure 1
Note: This action will display the dash symbol in the character field on the glass display.
(See Figure 2)
Figure 2
8.
Press the Confirm/Enter button.
(See Figure 3)
This action will display the “VIO 300D Setup” menu. The
unit’s date will be visible on the top of the display.