IMPORTANT!
Please read all information carefully. These instructions for use do
not replace the user manual of the smoke evacuation and
electrosurgical unit used! Read the user manual of each device and
ask your distributor in case of doubt!
1 Intended Use
The smoke evacuation pencil is intended for monopolar cutting and
coagulation, and for simultaneous smoke evacuation.
2 Normal use
The electrosurgical pencil is connected to the following device:
• Electrosurgical generators that have been tested to the IEC 60601
standard. Refer to the generator documentation and smoke
evacuation system to ensure compatibility.
3 Maximum electrical capacity
The maximum electrical capacity of this product is:
• 4500 Vp
4 Safety instructions
WARNING!
This product may be used only by trained medical staff, who have
been shown how to use it according to the instructions for use.
Check the product for damage before use, including all the insulation.
If damaged, do not use this product!
Do not use in the presence of combustible or explosive materials!
Never place the instrument on or in immediate proximity to the
patient. The use of a holster is recommended.
Protect this product from any form of mechanical damage! Do not
throw! Do not use force! If there is a cable, and unless stated
otherwise, do not kink it or wrap it around the accompanying
product!
-9-
Summary of Contents for 20321-040
Page 36: ...E0022 0936 01 111C 2 05 05 2020 20321 040 20321 041 20321 042 20321 043 33...
Page 37: ...CE 34...
Page 38: ...1 2 IEC 60601 3 4500 Vp 4 35...
Page 39: ...5 5 1 5 2 5 3 5 4 3 36...
Page 78: ...E0022 0936 01 111C 2 05 05 2020 20321 040 20321 041 20321 042 20321 043 75...
Page 79: ...76...
Page 80: ...1 2 IEC 60601 3 4500 Vp 4 77...
Page 81: ...5 5 1 5 2 5 3 78...
Page 82: ...5 4 3 5 5 6 10 40 C 30 75 70 106 7 25 55 C 25 70 C 10 100 70 106 79...
Page 89: ...E0022 0936 01 111C 2 2020 5 05 20321 040 20321 041 20321 042 20321 043 86...
Page 90: ...CE 87...
Page 91: ...1 2 ERBE IEC 60601 3 4500Vp 4 88...
Page 92: ...5 5 1 5 2 5 3 5 4 5 5 89...
Page 94: ...91 E0022 0936 01 111C 2 2020 05 05 20321 040 20321 041 20321 042 20321 043...
Page 95: ...Gebrauchsanweisung beachten CE 92...
Page 96: ...1 2 IEC 60601 3 4500Vp 4 93...
Page 97: ...5 5 1 5 2 5 3 5 4 HF 3 5 5 94...
Page 99: ......