PSI 9000 3U Series
Schematic diagram:
Application and result:
O
ffse
t
t
A
A
m
plit
ud
e
t1
t2
A triangular wave signal for output current (only effective
in current limiting) or output voltage is generated. The
positive and negative slope times can be set indepen
-
dently.
The offset shifts the signal on the Y-axis.
The sum of the intervals t1 and t2 gives the cycle time
and its reciprocal is the frequency.
Example: a frequency of 10 Hz is required and would
lead to periodic duration of 100 ms. This 100 ms can
be freely allocated to t1 and t2, e.g. 50 ms:50 ms (isos
-
celes triangle) or 99.9 ms:0.1 ms (right-angled triangle
or sawtooth).
3.9.7 Rectangular function
The following parameters can be configured for a rectangular wave function:
Value
Range
Description
I(A), U(A)
0...(Nominal value - (Off)) of U, I A = Amplitude of the signal to be generated
I(Off), U(Off) 0...(Nominal value - (A)) of U, I
Off = Offset, based on the foot of the rectangular wave
t1
0.1 ms...36000 s
Time (pulse width) of the upper level (amplitude)
t2
0.1 ms...36000 s
Time (pause width) of the lower level (offset)
Schematic diagram:
Application and result:
O
ffs
et
t
A
A
m
plit
ude
t1
t2
A rectangular or square wave signal for output current
(only effective in CC mode) or output voltage is gener
-
ated. The time values t1 and t2 define how long the
upper level of the pulse and how long the lower level of
the pause are effective.
The offset shifts the signal on the Y-axis.
Intervals t1 and t2 can be used to define a duty cycle. The
sum of t1 and t2 gives the cycle time and its reciprocal
is the frequency.
Example: a rectangular wave signal of 25 Hz and a
duty cycle of 80% are required. The sum of t1 and t2,
the period, is 1/25 Hz = 40 ms. For a duty cycle of 80%
the pulse time (t1) is 40 ms*0.8 = 32 ms and the pause
time (t2) is 8 ms.