13
U
SING
S
UPPLEMENT
AL
O
XY
GEN
INLET
WARNING
•
Use only medical grade oxygen.
•
Ensure that the device is ventilating before the
oxygen supply is turned on.
•
The oxygen flow must be turned off when the device
is not ventilating so that oxygen does not
accumulate within the device. The accumulation of
oxygen presents a fire risk.
1
2
3
5
4
6
Air inlet and pollen filter
AC / DC Power connector
STANDBY Button
O
2
input
FiO
2
/SpO
2
connector
Remote Alarm connector
1
2
3
4
5
6
•
Plug in the oxygen adaptor (supplied with the EO 150) to the oxygen inlet 4.
•
Attach the end of the oxygen supply tube to the oxygen adaptor.
•
Attach the other end of the oxygen supply tube to the oxygen source.
•
Start ventilation.
•
Turn on the oxygen and adjust for the prescribed flow rate or FiO
2
level.
•
Make sure that the oxygen source has been switched off before turning off
the ventilation.