CLM-194
User Guide
EOPTIS SRL
Via Alto Adige, 152 – 38121 Trento (TN) – ITALY
Tel.: +39 0461 260 552 – Fax: +39 0461 260 617
www.eoptis.com – [email protected]
Page 15 of 29
CLM-194
Rev. 1.05 - 06/2014
Copyright © 2013-2014 EOPTIS SRL
5.3.1
H
OW TO
P
ERFORM A
C
OLOUR
D
IFFERENCE
C
HECK
A
NALYSIS
1.
Select the Colour Difference Check function from the Operating Mode selector
2.
Set the Acceptance Thresholds
3.
Acquire a desired
Reference
sample. Either select an existing sample from the
References
list or let the
software detect the closest to
Measure
, belonging to the selected group of the
Reference
list.
4.
Acquire the data of the
Measure
sample under evaluation. If acquisition has already been performed, any
colour in the
Measure
list can be selected by double-clicking on it.
5.
The software performs the colour difference calculation and shows the results of acceptance threshold
validation and plots the (L*; a*; b*)
Measure
coordinates against the
Reference
ones.
5.3.2
U
SING THE
I
NTERFACE
The acquired colour is displayed on the monitor as a guide only. It is very likely that it will appear
to be different from the actual colour of the sample. This is due to several factors (such as the
ambient light, monitor calibration, gamma settings, …) which must be controlled and fine tuned in
order to for the colours to match.
The
Reference
can be selected in two different ways, using the
Reference
Mode selector ④.
Manual mode.
Manual mode is enabled by selecting the
MANUAL
option in the Reference Mode selector ④.
In Manual mode, the user selects a colour from the
Reference
List by double-clicking on it, by dragging it to the
Reference
colour display area ③, or by acquiring a new
Reference
colour. Manual mode allows the colour
difference of the current
Measure
(either that just acquired or selected from the
Measure
list) to be evaluated
with respect to a
Reference
which is fixed and does not change during operations.
Automatic mode
In automatic mode the software automatically chooses the
Reference
colour selecting the
most similar (lower ΔE*
ab
) to the current
Measure
. If the
ALL
option is selected in the
Reference
Mode selector
④, the colour will be chosen among all colors in the
Reference
list. If the corresponding group name is
selected, a colour group will be chosen. Automatic mode allows the user to highly speed up the Colour
Difference Check when a large set of samples comprised of a subset of various colours is to be checked. The
user does not need to select the right
Reference
when a new sample is being acquired since the software
automatically detects it. The automatic selection process is effective even on very similar colors if a group of
Reference
colours to be selected is specifically built and the ΔE*
ab
difference between these colors is larger
than the highest Acceptance Threshold.
The software displays the colour difference between
Measure
and
Reference
in terms of CIE 1976 a,b (CIELAB) ΔE*
ab
colour difference ⑤. The CIELAB lightness difference ΔL*, CIELAB chroma difference ΔC*
ab
, Δa* and Δb* difference
are also provided ⑥.
Δ
ab
√(
L
L
)
(
a
a
)
(
b
b
)
Δ
ab
√(
)
(
)
√(
)
(
)
ΔL L
L
;
Δa a
a
;
Δb b
b
being (L*
1
; a*
1
; b*
1
) the
Measure
coordinates and (L*
0
; a*
0
; b*
0
) the
Reference
coordinates.
The colour difference ΔE*
ab
is then compared to the user-defined Acceptance Thresholds ⑦ and if the corresponding
threshold is not exceeded, the
Measure
is ranked to one of the three acceptance classes (1, 2, 3). If the largest
threshold is exceeded, the
Measure
is ranked to the "void" (x) class.