14
4. Connect the testing points with the test pen.
5. Read the voltage displayed on the screen, it may include the numerical value, decimal point and cathode/anode
determined.
Voltage measuring menu:
REL Pattern
Data hold/Auto record
Pattern
AC/DC
Function
Waveform/Readout
Function
REL
H00/A-H
~/---
REL(Clear)
Store
AC/DC
Waveform/Readout
a. Press the “REL” key to enter REL pattern.
b. Press the “Storage” key to enter Data hold/store/Auto record pattern.
c. Press the “AC/DC” key to select the AC/DC function
d. Press the “Waveform/Readout” key to enter the waveform pattern; it will display the waveform of the current voltage or a
list of stored data and waveform.
AC/DC Current (mA, 20A) measuring
Warning!
• It’s strictly prohibit using the 20A input terminal to measure the voltage; please don’t operate a AC current
measuring in a circuit with voltage over AV 250V.
• Do not test over 15 second in every 15 minutes when using the 20A Current measuring range. Otherwise,
the meter and test pen connecting wire may be damaged.
AC/DC Voltage measuring illustration
Summary of Contents for ET201