ZHE+ Zero Headspace Extractor:
Operation and Instruction Manual
Maintenance and Warranty
+1.800.343.5319 or +1.843.881.6560 envexp.com
Environmental Express
• 4
Product Information:
Item
#
Date
of
Purchase
ZHE+ Zero Headspace Extractor Serial #
Please record the serial # of your ZHE+ zero headspace extractor here for easy reference. Your serial # is located on the back of the unit.
Maintenance and Care
The ZHE+ zero headspace extractor should be thoroughly cleaned and inspected
before each use. A non-abrasive laboratory cleanser such as Alconox® or Sparkleen®
is suitable. Do not use steel wool or abrasives such as Comet®. It should be rinsed
with laboratory-grade water. All O-rings should be inspected for signs of wear before
each use, and frequent replacement of O-rings (Item # ZVR1000), along with the rings
within the plug valve (Item # ZVR9903), is recommended to reduce both sample and
blank contamination.
Note:
316 grade stainless steel is a hard, inert metal, but possible to damage.
Avoid rough handling and overexposure to corrosives.
Periodically a leak test on the ZHE+ unit should be performed to ensure proper
operation. To perform a leak test, fully assemble the unit and pressurize to 50 psi.
Leave pressurized for an hour and check the gauge for any pressure loss. If a
pressure loss is found, re-pressurize the unit, place under water and look for bubbles.
It may be necessary to replace O-rings or tighten fittings.
Warranty
Environmental Express, Inc. warrants materials and workmanship on the ZHE+ zero
headspace extractor for the period of one year from the date of shipment. This
warranty extends to parts and labor, applies only to damage or failure caused by
normal laboratory use and is limited to product repair.
After proper notification, any defective ZHE+ zero headspace extractor or part thereof
may be returned to Environmental Express, Inc. for, at Environmental Express’ option,
replacement or refund. In no event shall Environmental Express be liable for any
incidental, consequential, special, or indirect damages of any kind pursuant to the
purchase and/or use of this product.
Call +1.800.343.5319 or +1.843.881.6560 envexp.com
2345 A Charleston Regional Pkwy • Charleston, SC 29492