![Environmental Express 99561-84 Instruction Manual Download Page 4](http://html1.mh-extra.com/html/environmental-express/99561-84/99561-84_instruction-manual_2416080004.webp)
COD 200 Instruction Manual
4
Instrument
automatic wavelength selection,
direct reading colorimeter
Light source
LEDs, interference filters (IF) and photo
-
sensor in transparent
cell chamber
Photometric Range
-2600 - 2600 mAbs
Operating Wavelenght
430 nm Δλ = 5 nm
610 nm Δλ = 6 nm
Wavelength accuracy
± 1 nm
Photometric accuracy*
3 % FS (T = 20 °C – 25 °C)
Photometric resolution
0.01 A
Auto-OFF
automatic switch off 25 minutes after last
keypress (different reaction times depend
-
ing on available method)
Display
backlit LCD (on keypress)
Time
real time clock and date
Calibration
user and factory calibration
resetting to factory calibration possible
Ambient conditions
temperature: 5– 40 °C
rel. humidity: 30–90 % (non-condensing)
Waterproof
floating; as defined in IP 68 (1 hour at 0.1
meter)
CE
Certificate for Declaration of CE-Conformity
available for download at:
*measured with standard solutions
Technical Specifications
Summary of Contents for 99561-84
Page 5: ...V1_04 2022 5 Guidelines for colorimetric measurements ...
Page 9: ...V1_04 2022 9 ...
Page 10: ...COD 200 Instruction Manual 10 ...
Page 15: ...V1_04 2022 15 ...
Page 16: ...COD 200 Instruction Manual 16 ...
Page 21: ...V1_04 2022 21 ...
Page 22: ...COD 200 Instruction Manual 22 ...
Page 27: ...V1_04 2022 27 ...
Page 28: ...COD 200 Instruction Manual 28 ...
Page 33: ...V1_04 2022 33 ...
Page 34: ...COD 200 Instruction Manual 34 ...
Page 35: ...V1_04 2022 35 ...
Page 36: ...COD 200 Instruction Manual 36 ...
Page 37: ...V1_04 2022 37 ...
Page 38: ......
Page 39: ......
Page 40: ...Environmental Express www environmentalexpress com ...