Measuring
1. Select the method on the device: HR
Prepare two reaction vials.
Mark one as a blank.
Put
0.2 ml deionised water
in the blank
Put
0.2 ml sample
in the
sample vial.
Close vials.
Carefully invert several
times to mix the contents.
Note: Will get hot!
Remove the vial from the
thermoreactor.
(Note: vial
will be hot!)
Allow vial(s) to cool to 60
°C.
Invert several times to mix
the contents.
Seal the vials in the pre-
heated thermoreactor for
120 minutes at 150°C
COD HR / 200 - 15000 mg/l / HR