COD 200 Instruction Manual
24
COD MR / 20 - 1500 mg/l / MR
Sampling
1. In the standard version, chloride interferes from a concentration of 1000 mg/L. In the
mercury-free version, the disturbance depends on the chloride concentration and the
COD. Concentrations from 100 mg/L chloride can lead to significant disturbances
here. See "Sample Preparation" Section for notes on removing chloride interfernce.
2. Suspended solids in the vial can lead to incorrect measurements and so to avoid
this, it is important to place the vials carefully in the sample chamber as the method
necessitates a build-up of precipitate at the bottom of the vial.
3. The outer walls of the vial must be clean and dry before the analysis is carried out.
Fingerprints or water droplets on the vial lead to incorrect measurements.
Notes About Method
1. The blank is stable when stored in the dark.
2. Blanks and test vials must be from the same batch.
3. Do not place hot vials in the sample chamber. The most stable measured values can
be determined if the vials are left standing overnight.
Summary of Contents for 99561-84
Page 5: ...V1_04 2022 5 Guidelines for colorimetric measurements ...
Page 9: ...V1_04 2022 9 ...
Page 10: ...COD 200 Instruction Manual 10 ...
Page 15: ...V1_04 2022 15 ...
Page 16: ...COD 200 Instruction Manual 16 ...
Page 21: ...V1_04 2022 21 ...
Page 22: ...COD 200 Instruction Manual 22 ...
Page 27: ...V1_04 2022 27 ...
Page 28: ...COD 200 Instruction Manual 28 ...
Page 33: ...V1_04 2022 33 ...
Page 34: ...COD 200 Instruction Manual 34 ...
Page 35: ...V1_04 2022 35 ...
Page 36: ...COD 200 Instruction Manual 36 ...
Page 37: ...V1_04 2022 37 ...
Page 38: ......
Page 39: ......
Page 40: ...Environmental Express www environmentalexpress com ...