Measuring Free Chlorine with Powder Packs
1. Select the method on the device: CL2
Fill 24 mm vial with
10 ml
sample.
Close vial
Place
sample vial
in the
sample chamber. Pay
attention to the positioning.
Press the
ZERO
button
Remove the vial from the
sample chamber.
Add
Chlorine FREE-DPD/
F10
powder pack
Close vial
Invert several times to mix
the contents (20 sec.)
Chlorine PP / 0.02 - 2 mg/l / CL2