MAC 10  Original VE5 SERVICE MANUAL 

 

PAGE 12 

 

 

Note : When turning the unit “ON” from the “OFF” position of the speed control, the fan 
is at the highest speed. Turning the speed control knob clockwise will lower the airflow. 

VII.  Trouble Shooting: 

 

Low Air Velocity

  

                           

Step 1. 

Check prefilter media; replace or clean as necessary. 

 

Step 2. 

Flip switch from “Low” to “Medium” or “Medium” to “High” on units with 3-speed 
switch. 

 

Step 3. 

Adjust variable speed control (only units with 

optional

 Speed Control) for 

higher blower output. 

 

Step 4. 

Check power supply for proper voltage, amperage and distribution frequency. 

 

Step 5. 

Replace HEPA/ULPA filter if the air velocity remains low. 

 
 

High Air Velocity

 

 
Step 1. 

Flip switch from “HIGH” to “MEDUIM” to “LOW” on units with 3-speed switch.

 

 

Step 2. 

Adjust variable speed control (only units with 

optional

 Speed Control) for lower 

blower output. 

 
 
 

Non-Laminar Flow and/or Excessive Contamination

 

 

Step 1. 

Insure that no large obstructions are upstream of airflow pattern. 

 

Step 2. 

Determine that no other air-moving devices are operating in or around clean 
room which disrupt room’s airflow pattern. 

 

Step 3. 

Check air velocity and if low, conduct the “Low Air Velocity” procedure outlined 
above. 

 

Step 4. 

Conduct smoke and photometer test on HEPA filter. Seal or replace HEPA 
filter as necessary. 

 

 
 

Summary of Contents for MAC 10 VE5 RSR

Page 1: ...E5 FAN FILTER MODULE INSTALLATION AND SERVICE MANUAL STANDARD RSR RSRE MODELS Mac 10 and VE5 are registered Trademarks of the Envirco Corporation Albuquerque New Mexico U S A US Patents 4 560 395 and...

Page 2: ...rer If you have any questions contact the manufacturer ENVIRCO CORPORATION FEDDERS INTERNATIONAL Envirco Office Headquarters 5601 Balloon Fiesta Parkway N E No 728 Fuzhou Road Albuquerque New Mexico 8...

Page 3: ...II SERVICE REMOVAL AND INSTALLATION OF THE ROOM SIDE REPLACEABLE FILTER RSR RSRE 5 IV SERVICE REMOVAL AND INSTALLATION OF THE MOTOR STANDARD AND RSR MODELS 7 V SERVICE REMOVAL AND INSTALLATION OF THE...

Page 4: ...prefilter is recommended every three to six months Step 1 To gain access to the prefilter remove the ceiling panel next to the unit if applicable Step 2 Switch the ON OFF switch to the off position S...

Page 5: ...lding the HEPA ULPA filter to the lid assembly Step 3 Lift the lid assembly off the HEPA ULPA filter see Figure 2 Discard the used filter as per requirements of the applicable regulations Note Before...

Page 6: ...ve the six screws and clips remove filter blocks and lower the filter as shown in Figure 3 Step 4 Inspect filter for visible damage if damaged set aside for replacement or repair Step 5 Place the filt...

Page 7: ...MAC 10 Original VE5 SERVICE MANUAL PAGE 6 Figure 4 4 1 2 3 5 6...

Page 8: ...t off cover Step 5 Make note of all wire locations for reinstallation later Step 6 Disconnect the two brown wires from the capacitor using a pair of pliers Step 7 Disconnect the motor wiring from the...

Page 9: ...MAC 10 Original VE5 SERVICE MANUAL PAGE 8 Figure 5 Disconnect motor leads from capacitor ON OFF switch and three speed switch or speed control Prefilter Motor Venturi Ring Blower Wheel...

Page 10: ...hile supporting the baffle assembly from below remove the four corner screws on the baffle assembly and lower the assembly see Figure 6 Step 3 Prior to removing motor blower assembly remove electrical...

Page 11: ...oval of the motor mount bracket measure the precise location of the bracket on the motor Remove the bracket Step 7 Replace with the new motor and reassemble by reversing the above steps Set the locati...

Page 12: ...ould be conducted to determine the filter condition and appropriate fan speed setting Figure 8 Optional Speed Control Units furnished with the optional speed control the speed control is standard with...

Page 13: ...oltage amperage and distribution frequency Step 5 Replace HEPA ULPA filter if the air velocity remains low High Air Velocity Step 1 Flip switch from HIGH to MEDUIM to LOW on units with 3 speed switch...

Page 14: ...10 Original VE5 SERVICE MANUAL PAGE 13 VIII MAC 10 VE5 Wiring Diagrams Figure 10 Figure 11 Standard Wiring for MAC 10 ON OFF switch Speed Control 10 Capacitor Optional Wiring for MAC 10 With Speed Con...

Page 15: ...Standard Filter HEPA ULPA Optional 2 x 2 1 69514 004 69514 018 2 x 3 1 69514 019 69514 012 2 x 3 5 1 69514 020 69514 021 2 x 4 1 69514 006 69514 007 HEPA Filter RSR 2 x 2 1 69391 005 69391 006 2 x 3 1...

Page 16: ...ny other commercial loss This warranty excludes certain expendable items such as light tubes prefilters etc ENVIRCO expressly disclaims and excludes from this warranty any responsibility for equipment...

Page 17: ...MANUAL PAGE 16 cleanroom start up may be time consuming and recommends using 3 Velgrid readings taken on a diagonal as shown below Recommended Testing 8 readings with a Velgrid Factory Approved Testin...

Reviews: