MAC 10 Original VE5 SERVICE MANUAL
PAGE 9
V.
SERVICE: Removal and Installation of the RSRE Motor
WARNING!
Disconnect the unit from the electrical power source before attempting any
service.
WARNING!
Electrical service should be performed by licensed electricians or authorized
ENVIRCO service technicians.
Step 1.
To gain access to the motor, remove the gel seal filter (see Section III).
Step 2.
While supporting the baffle assembly from below, remove the four corner
screws on the baffle assembly and lower the assembly.(see Figure 6)
Step 3.
Prior to removing motor/blower assembly, remove electrical box cover (located
underneath the lid panel) to expose motor connectors. Disconnect the power.
Step 4.
While supporting the motor/blower assembly from below, remove the eight
screws on the motor plate and lower the motor/blower assembly.(see Figure 7)
Figure 6
Gel Seal Filter
Lid Panel
Motor Plate
(Motor/Blower Assy)
Electrical connectors located
on inner prefilter wall
Plenum
Baffle Assembly