6 Basic System Setup
VIEW 370/373
24
(PC-000855-MA Issue 6, December 2021)
ENVEA
UK Ltd
6.4
Initial System Checks
6.4.1
On Power-up
Switch on the external power supply to the control unit. Check that the Power LED
is ON, and the graphical display is illuminated. When using the control unit for the
first time, communication must be established between the control unit and all
connected devices in the network. If the control unit is being reconfigured, then
all existing settings can be cleared by performing a Master Reset (see below for
details).
6.4.2
Performing a Master Reset
CAUTION
Loss of data- Performing a Master Reset deletes all settings and logged
data.
Download the data to a PC using the ENVEA software tool before carrying
out a system reset.
1.
Open the
Setup
menu screen.
2.
Select
Other Functions | Reset Functions
.
3.
Select
Master Reset
to erase the current configuration.
6.4.3
Checking the Sensor Comms Mode
The communications (Comms) mode is set by ENVEA prior to installation and
should be checked during the system installation.
1.
Select
Setup | Other Functions | Advanced Functions
.
Check the
Sensor Comms Mode
. The sensor communication settings
should match (refer to
on page 18 for details). If there are
problems, check that the sensor communications setting (SW101) is the
same as the Sensor Comms Mode settings in the control unit.
Sensor comms
Mode
Baud rate
Parity
Standard
ASCII
19200
Odd
Optional
RTU
19200
None