O
PERATION
Combination Blanket/Fluid Warmer
▪
Operator’s Manual
▪
M N-39940
▪
Rev 3
▪
8/20
30
How to Operate the Blanket Warmer
(Bottom Cavity)
Before you begin
The warmer must be connected to electric power and turned on.
Procedure
To operate the warmer, do the following.
Step
Action
1.
Set
the temperature using the arrow buttons
.
2.
Open
the door and load the dry, cotton blankets into the warmer.
Close
the door.
During the warming
process
3.
Press
the temperature recall button
to view the measured cavity
temperature.
The measured cavity temperature will display for five seconds. Then, the
set-point temperature will display.
4.
Rotate
the blankets daily from the bottom to the top of the load to ensure
equal heating and to prevent discoloration.
Result
The blankets are now warming.
WARNING:
Personal injury hazard.
Do not operate the warmer in the presence of a flammable
anesthetic mixture with air, oxygen, or nitrous oxide; in
oxygen-enriched environments; or in any other potentially
explosive environment.
WARNING:
Personal injury hazard.
Make sure the blanket support assembly and shelves are
installed to prevent the blankets from being scorched or
discolored.
NOTICE
Do not overload the cavity.
Allow 1” (25mm) between the blankets and the interior walls.
NOTE:
The temperature set-point range is 90°F – 160°F
(32°C – 71°C).
NOTICE
Do not warm items containing plastic, rubber, metal
snaps, studs, hooks, etc.